baño de temperatura portable Fluke 7109A

Nuevo producto

Calibrate more sanitary sensors at a time – with confidence

The 6109A and 7109A baths calibrate most temperature sensor applications used in clean process manufacturing, including pharmaceutical bioreactors (-10 C to 100 °C), chemical reactors (200 °C), steam-in-place process (122 °C to 140 °C), autoclave sterilization (120 °C to 135 °C), food storage freezers (-25 °C), and food processing (0 °C to 220 °C).

Más detalles

Más

The 6109A and 7109A baths calibrate most temperature sensor applications used in clean process manufacturing, including pharmaceutical bioreactors (-10 C to 100 °C), chemical reactors (200 °C), steam-in-place process (122 °C to 140 °C), autoclave sterilization (120 °C to 135 °C), food storage freezers (-25 °C), and food processing (0 °C to 220 °C).

A large tank volume (112 mm square tank opening x 154 mm depth) lets you immerse up to four sanitary sensors at the same time. Calibrate a batch of odd shaped temperature sensors of varying lengths and diameters, and still have room for a reference thermometer. Fast heating and cooling times let you get the job done without spending a lot of time waiting for the bath to come to temperature and stabilize. Increase your sanitary sensor calibration throughput and decrease expensive production downtime.

Excellent system display accuracy of ± 0.1 °C. Accuracy covers all sources of error including calibration uncertainty, stability, uniformity, and repeatability. NVLAP accredited calibration is included standard. Fluke Calibration’s high metrological standards and conservative specifications mean you can have confidence in the measurements you make with the 6109A and 7109A Portable Calibration Baths.
 

Summary Specifications

General specifications
SizeHeight: 382 mm (15 in)
Width: 242 mm (9.5 in)
Depth: 400 mm (15.7 in)
Weight6109A: 16 kg (35 lb)
7109A: 20 kg (44 lb)
Fluid volume2.5 liters (0.66 gallons)
Fluid working area75 mm x 75 mm (3 in x 3 in)
Maximum fluid depth154 mm (6.1 in)
Factory calibrationTraceable NVLAP-accredited calibration included

Temperature specifications
The temperature specifications describe the Absolute Instrumental Uncertainty at 95 % level of confidence (coverage factor k = 2) within one year from calibration. The specifications include environmental temperature effects from 13 °C to 33 °C.
 6109A7109A
Range (at 25 °C ambient temperature)35 °C to 250 °C–25 °C to 140 °C (–15 °C cover off)
Accuracy (maximum permissible error)0.1 °C0.1 °C
Display resolution0.01 °, 0.001°0.01 °, 0.001°
Stability0.015 °C0.01 °C
Typical uniformity0.03 °C up to 200 °C
0.04 °C above 200 °C
0.02 °C
Repeatability0.04 °C0.04 °C
Typical heating time35 °C to 100 °C: 25 minutes
100 °C to 250 °C: 45 minutes
–25 °C to 25 °C: 35 minutes
25 °C to 140 °C: 55 minutes
Typical cooling time250 °C to 100 °C: 35 minutes
100 °C to 35 °C: 55 minutes
140 °C to 25 °C: 45 minutes
25 °C to –25 °C: 75 minutes
Typical settling time15 minutes10 minutes

Process Input Module specifications (-P models)
The Input Module specifications describe the Absolute Instrumental Uncertainty at 95 % level of confidence (coverage factor k = 2) within one year from calibration. The specifications include calibration uncertainty, linearity, repeatability, resolution, stability, and environmental temperature effects from 13 °C to 33 °C.
FunctionRangeAccuracy (Maximum Permissible Error)
Reference input resistance0 Ω to 42 Ω
42 Ω to 400 Ω
0.0025 Ω
0.006 %
Reference input temperature-25 °C to 250 °C0.007 % + 0.015 °C
Resistance sensing current1 mA8 %
DUT 4-wire resistance0 Ω to 31 Ω
31 Ω to 400 Ω
0.0025 Ω
0.008%
DUT 3-wire resistance accuracy0 Ω to 400 Ω0.12 Ω
Thermocouple mV-10 mV to 100 mV0.025 % + 0.01 mV
Reference junction temperature0 °C to 40 °C0.35 °C
Thermocouple temperature-25 °C to 250 °CJ: 0.44 °C
E: 0.44 °C
L: 0.42 °C
K: 0.49 °C
N: 0.57 °C
U: 0.48 °C
T: 0.53 °C
M: 0.48 °C
0 °C to 250 °CR: 1.92 °C
D: 1.12 °C
S: 1.88 °C
G/W: 3.34 °C
C: 0.84 °C
mA range0 mA to 22 mA0.02 % + 0.002 mA
Loop power24 V dc± 6 V

Reseñas

Escribe tu opinión

baño de temperatura portable Fluke 7109A

baño de temperatura portable Fluke 7109A

Calibrate more sanitary sensors at a time – with confidence

The 6109A and 7109A baths calibrate most temperature sensor applications used in clean process manufacturing, including pharmaceutical bioreactors (-10 C to 100 °C), chemical reactors (200 °C), steam-in-place process (122 °C to 140 °C), autoclave sterilization (120 °C to 135 °C), food storage freezers (-25 °C), and food processing (0 °C to 220 °C).